Purisys CPF201
Brand: Endress+HauserPurisys CPF201
Brand: Endress+HauserSpecs at a Glance
Process temperature:
2°C - 75°C
(36°F - 167°F)
(36°F - 167°F)
Process pressure:
max. 3 bar
(43.5 psi)
(43.5 psi)
Measuring range:
pH 1-13 (<100µS/cm)
Field of application
Purisys CPF201 is the specialist for pure and ultrapure water. Its integrated stainless steel assembly provides stable and flow-independent measurement at lowest conductivities. Purisys ensures optimum water quality and guarantees highest safety of your products or assets.
Boiler feed water
Power plant applications
Pharmaceutical water for injection
Key Benefits
- Stable pH reading and response
- Elimination of flow dependency errors
- Patented porous PTFE reference junction
- Easy installation
Specifications
Brand
Industry
- Power & Energy
- Life Science
Product Family
- Purisys
pH
Application
- Ultra pure water (<100µS/cm); power plants
Characteristic
- Bypass measurement system made of stainless steel with gel electrode and salt storrage for measurements in ultra pure water.
Measuring range
- pH 1-13 (<100µS/cm)
Measuring principle
- Glass-gel compact electrode with PTFE diaphragm and salt (KCl) storrage
Design
- Electrode in stainless steel boundary surface installed in a stainless steel flow assembly for potential free measurements in ultrapure water with lowest conductivity.
Material
- Glass, stainless steel
Dimension
- Installation on a plate (305 mm x 305 mm).
- (11.89 inch x 11.89 inch)
Process temperature
- 2°C - 75°C
- (36°F - 167°F)
Process pressure
- max. 3 bar
- (43.5 psi)
Temperature sensor
- Optional with integrated Pt100 oder Pt1000.
Connection
- TOP68 connection head or fixed cable connection.
Ingress protection
- IP68
- Ultra Pure Water (<100 µS/cm)
- Power Plants